vH2O2 Vapor Concentration & Condensation Point

蒸发过氧化氢PPM对冷凝的影响
乔尼·帕坦森(Joni Partanen)
Product Manager
Published: 2022年2月3日
Life Science

该博客是四部分系列中的第四个博客,我们描述了过程参数如何影响蒸发的过氧化氢生物污染应用。

第一个博客:Humidity, condensation point, and maximum achievable vH2o2

Second blog:H2o2Solution Concentration Ratio, Condensation Point & Maximum Achievable vH2o2

第三博客:温度,冷凝点和最大可实现的VH2o2PPM在生物污染中

In this series, we propose four basic process parameter rules. This blog focuses on the 4th rule:

增加h2o2蒸气浓度降低了空气可以容纳的水蒸气量,因此,凝结会更早。

在图7中(来自我们的白皮书), we can see that increasing the H2o2蒸气浓度降低了空气可以容纳的水蒸气量,并更早发生凝结。

图片
Figure 7. vH2O2 condensation points at given temperatures and ppm vH2O2 (at each point RS = 100 %, maximum %RH varies according to curves)

Each point in Figure 7 is a condensation point, meaning thatRelative Saturationis 100 %. Temperature is on the x axis and ppm vH2o2在y轴上。图中的曲线显示在给定温度和PPM VH浓度下的最大相对湿度2o2

如图所示,在20°C和300 ppm的过氧化氢下,相对湿度将为60%,相对饱和度为100%。如果我们增加VH2o2to 600 ppm at 20 °C, relative humidity is then 39 % and relative saturation is 100 %.

通过将空气温度提高到40°C,VH2o2浓度在300 ppm时,相对湿度将为87%,相对饱和度为100%。温度越高,空气可以容纳的水蒸气越多;相对湿度增加。

To summarize our proposed rules for controlling condensation in vaporized hydrogen peroxide bio-decontamination applications:

  • Lowering the initial level of humidity increases the amount of H2o2凝结之前可以使用的蒸气。
  • Increasing the temperature increases the amount of water and hydrogen peroxide vapor the air can hold, thereby increasing the maximum achievable ppm vH2o2
  • 较高的浓度h2o2solution increases the H2o2凝结之前可以使用的蒸气。
  • 增加h2o2蒸气浓度降低了空气可以容纳的水蒸气量,因此,凝结发生更快

Conclusion

over this series of blogs (based on our白皮书), we have demonstrated how an in-depth understanding of the relation between critical process parameters allows for the development of effective, repeatable vH2o2生物污染周期。

Single-parameter measurement is often insufficient for monitoring and ineffective for process control. We have also shown why relative saturation is an important value in accurately predicting condensation. This is why Vaisala’s uniquePeroxcap®技术measures multiple parameters in a single sensing unit, including: hydrogen peroxide vapor ppm, temperature, dew point, vapor pressure, and humidity as both relative humidity and relative saturation.

请在下面的字段中发表评论,或联系我们如果你有任何问题!

按需网络研讨会:
Common vH2O2 bio-decontamination challenges and how to solve them

在本网络研讨会中,我们回答了我们在生物污染中收到的几个关于蒸发过氧化氢传感器的最常见问题,包括:

  • How to integrate measurements into your system: Connectivity - analog, digital, and host devices
  • 传感器放置的最佳实践
  • ppm不会增加...为什么不呢?
  • Relative humidity and relative saturation: What's the difference?
  • 材料在环境中的影响
  • 如何使用测量数据
  • Analyzing the environment
  • 了解您的过程