blog

H2O2吸收,传感器上的催化层,蒸气注射速率等等…

Vaporized Hydrogen Peroxide sensor
生命科学

Questions & Answers from our Recent Webinar on Hydrogen Peroxide Bio-decontamination

If you were unable to attend this webinar, you can随时观看。这是我们在网络研讨会期间无法解决的问题的答案。

快速调查 - 如何测量过氧化氢?

Please answer these 4 questions on the technology you currently use and what sort of ppm concentrations you measure.

网络研讨会问题和答案:


问题:
Do you have solution for testing H2o2吸收在材料中?


回答:

当我们开发HPP270系列(HPP271和HPP272)探针,我们测试了几种材料,仅选择不吸收或分解H2O2的探针材料。HPP271和HPP272探针用于高级和过程级别的测量,因此不是用于测试H2o2这可能会吸收客户的材料(流出研究)。

问题:
Can you explain how pressure and temperature vary during a bio-decontamination cycle?

回答:

slide presentation,请参考幻灯片13,该幻灯片显示隔离器中典型的非敏感性生物污染过程。
蒸发过氧化氢生物污染
如果您使用热空气和干燥空气降低湿度,温度往往会升高。同样,在调节阶段和生物污染阶段,温度升高,将过氧化氢蒸气注入到腔室中。在生物污染阶段,压力保持在相同的水平,但通常略高于隔离器的周围区域。

问题:为什么没有催化层的其他RH传感器显示出更高的值?这些传感器与没有催化剂层的传感器相同,不是吗?


回答:

过氧库传感器变暖,防止冷凝传感器在传感器上形成。即使在冷凝条件下,这也提供了可靠的测量。此外,传感器还具有化学清除功能,该功能有助于保持挑战H的准确性2o2环境。这purging process involves rapid heating of the sensor to remove possible contamination. The HPP272 probe provides two different humidity values: Relative Humidity (RH) and Relative Saturation (RS).

相对湿度是从催化版本的直接测量腐殖质。但是,相对饱和是从相对湿度h的计算参数H2o2和温度。据我们所知,没有其他其他过氧化氢传感器制造商具有这种可重复且可靠的参数,该参数表明包括过氧化氢蒸气的空气混合物的实际冷凝水平。

问题:
在蒸发H期间,相对饱和度和相对湿度如何与PPM水平的饱和相关2o2(35%)在包含区域?


回答:

您可以使用不同的H2o2液体生成H2o2蒸气,例如使用热板。h越高2o2液态氧化液的浓度,水蒸气较少。因此,您能够注入更多H2o2冷凝之前进入腔室。生成的过氧化氢蒸气PPM水平和温度会影响湿度水平。

我们将您推荐给推介会slides, 30 through 43。这些幻灯片显示了不同的温度和H2o2蒸气浓度会影响不同的湿度水平(相对湿度和相对饱和)。
vaporized hydrogen peroxide concentration by temperature graph

推介会。- 幻灯片30 - 43.此图显示了不同H中的RH%值2o2ppm浓度和温度值。

问题:
您能说明有关Vaisala氢过氧化氢传感器中使用的材料的更多信息吗?在这里指幻灯片50?

回答:

Different manufacturers' sensors contain different materials. As we learned today in our webinar, the selection of materials proximal to sensing elements is tremendously important to achieve accurate and repeatable measurements. During our Research & Development phase, Vaisala heavily tested all materials used in our measurement probesHPP271和HPP272。只有不吸收或分解H的材料2o2选择了蒸气,例如不锈钢和特氟龙。

This is crucial from the standpoint of the user; especially if isolators or rooms to be bio-decontaminated contain many materials that absorb, desorb or decompose H2o2汽。如果是这样,实现稳定的H2o2蒸气条件具有挑战性,并且由于发生混乱现象,曝气阶段可能会延长。

问题:
为什么湿度传感器包括催化过滤器?


回答:

根据该定义,相对湿度是水蒸气的量度。HPP270探针的催化层的功能是分解H2o2进入水和氧气。因此,只有水蒸气才能到达催化层下方的湿度传感器。

Vaporized hydrogen peroxide is an extremely harsh chemical that causes corrosion to many materials, especially if the process includes condensation. That’s why several hydrogen peroxide sensor manufacturers have included a catalytic filter or layer in their humidity sensors to protect the sensing element. Once H2o2is catalyzed, the sensor indicates the humidity only caused by water vapor pressure. Humidity sensors without catalytic filter drift quickly in environments with H2o2蒸气,因此没有提供可靠的测量结果。

问题:
我目前仅用于监测过氧化氢测量。您是否知道人们是否通常使用过氧化氢测量来控制过氧化氢的注射速率?


回答:

今天,大多数用户正在使用H2o2PPM值仅用于监视目的。但是,您可以使用可靠且可重复的测量来控制H2o2注射率使H2o2在生物污染阶段,PPM水平保持稳定。

另外,相对饱和度测量可用于控制和避免在需要时凝结。已经有一些公司使用测量h2o2ppm value to control H2o2生物污染过程中的注射率。我们已经看到公司越来越多地使用H2o2用于控制目的的蒸气浓度和相对饱和值。

问题:
我需要一直监测湿度 - 不仅在生物污染过程中。我可以使用Vaisala HPP探针测量生物污染周期之间的湿度吗?


回答:

HPP272探针是生物污染过程中湿度测量的好选择,也是常规生产过程中的湿度。HPP272相对饱和值向您显示何时在生物污染阶段空气混合物开始凝结时。当没有H2o2,%rs等于%RH。这意味着在正常生产过程中,您可以使用RH或RS参数进行湿度测量和控制。

这HPP272 probe is specifically designed to be used in highly condensing H2o2环境。因此,它提供可靠且可重复的测量数据(RS,RH,H2o2ppm和t)不仅在生物污染过程中,而且在生产过程中。

H2o2

评论

Periyasamy

7月9日,2021年
We need measure h202
And we need h2o2 convert in to water plus oxygen

珍妮丝·贝内特(Janice Bennett)

Jul 13, 2021
尊敬的先生,
您的评论已转发给我们的销售团队,他们将与您联系。感谢您的查询!